Z-1231-2019 Class I Terminated
FDA device recall Z-1231-2019 was initiated by OriGen Biomedical, Inc. on Jan 17, 2017 and is designated Class I. Reason for recall: Clamping the clear tube very close to the blue hub may cause a localized adhesive failure in the bond between the clear tube and the molded hub. Forcefully rotating the clamp while it is in place may… The recall status is terminated (terminated Jul 27, 2021). Affected quantity: 61.
Recall details
- Product Type
- Devices
- Report Date
- Jun 12, 2019
- Initiation Date
- Jan 17, 2017
- Termination Date
- Jul 27, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 61
Product description
Reason for recall
Distribution pattern
US: NY, AB, AR, FL, MO, NC, GA, LA, CO, AL, MA, CA, MI, WI, TX, DC, OH, CT, TN, IN, OR, IL, MN, QC, SC, KY, OK, AZ, UT, RI, NJ, SD, PA, MS, NM. Foreign (OUS): UAE, Saudi Arabia, Colombia, UK, Netherlands, Italy, Czech Republic, Germany, India, Japan, Canada, Thailand, Sweden, Belgium, Spain, Poland, South Africa, Chile