Z-1238-2015 Class II Terminated
FDA device recall Z-1238-2015 was initiated by Elekta, Inc. on Jan 29, 2015 and is designated Class II. Reason for recall: It is possible to position the Treatment Table with errors greater than 5mm and no inhibits are displayed, if the positional sensors are not correctly installed during corrective maintenance. Clinical… The recall status is terminated (terminated Nov 5, 2021). Affected quantity: 529 US 2951 Worldwide.
Recall details
- Product Type
- Devices
- Report Date
- Mar 25, 2015
- Initiation Date
- Jan 29, 2015
- Termination Date
- Nov 5, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 529 US 2951 Worldwide
Product description
Reason for recall
Distribution pattern
Worldwide Distribution - USA (nationwide) Distribution and to the countries of Albania, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Finland, France, Germany, Greece, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Lebanon, Libya, Lithuania, Malaysia, Malta, Mexico, Morocco, Myanmar, Namibia, Netherlands, New Zealand, Nigeria, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland and Syria.
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1238-2015
- openFDA file date
- Nov 10, 2026
- Loaded by RegDataLab
- Nov 11, 2026