Z-1246-2017 Class II Terminated

Recalled by Merge Healthcare, Inc. — Hartland, WI

FDA device recall Z-1246-2017 was initiated by Merge Healthcare, Inc. on Mar 28, 2016 and is designated Class II. Reason for recall: Eye Station images were not importing properly and were imported under "unknown" due to an issue when validating patients using only an Medical Record Number (MRN) The recall status is terminated (terminated Apr 17, 2020). Affected quantity: 1,627 sites potentially have the affected versions.

Recall details

Product Type
Devices
Report Date
Mar 1, 2017
Initiation Date
Mar 28, 2016
Termination Date
Apr 17, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,627 sites potentially have the affected versions

Product description

Merge Eye Station Import Utility (ESIU) when used with Merge Eye Station and Merge Eye Care PACS. The firm name on the labeling is Merge Healthcare, Hartland, WI.

Reason for recall

Eye Station images were not importing properly and were imported under "unknown" due to an issue when validating patients using only an Medical Record Number (MRN)

Distribution pattern

Distribution was nationwide to medical facilities. There was also government and foreign distribution, including Canada.

Code information

Merge ESIU: Versions 1.0, 1.1.1, 1.1.2, and 1.1.3; Merge Eye Station: All versions; Merge Eye Care PACS: Versions 4.1, 4.2, 4.3, 5.0, and 5.1

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1246-2017
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1246-2017. Report a data issue: hello@regdatalab.com.