Z-1262-2019 Class II Terminated

Recalled by GE Healthcare, LLC — Waukesha, WI

FDA device recall Z-1262-2019 was initiated by GE Healthcare, LLC on Mar 20, 2019 and is designated Class II. Reason for recall: During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector… The recall status is terminated (terminated Oct 5, 2020). Affected quantity: 31 units.

Recall details

Product Type
Devices
Report Date
May 8, 2019
Initiation Date
Mar 20, 2019
Termination Date
Oct 5, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
31 units

Product description

NM/CT 850 Product Usage: The GE NM/CT 850 system is intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body. It includes a general purpose Nuclear Medicine system using a variety of scanning modes supported by various acquisition types, and a Computed Tomography component which is intended specifically for enabling attenuation correction and anatomical localization of SPECT images.

Reason for recall

During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.

Distribution pattern

US and Austria, Canada, Croatia, France, Germany, India, Italy, Japan, Republic of Korea, Netherlands, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, UK

Code information

Mfg. Lot or Serial # System ID Model, Catalogue or Code # 85EX67018 4740081NM11 H3907AD 85EX67001 203789NM850 H3907AD 850X61005 319768NMCT H3907AD 850Y61006 319768NMH988 H3907AD 85EX67022 847618NM H3907AD 85EX67019 773975NMCT H3907AD 85EX67008 779696NM850 H3907AD 85EX67017 318212NM H3907AD 85EX67012 781729NM H3907AD 85EX67015 SHIPPED (AO 1623723) H3907AD 85EX67020 SHIPPED (GON 4745963) H3907AD 85EX67023 SHIPPED (GON 4704278) H3907AD 85EX67024 SHIPPED (GON 4710759) H3907AD 85EY67025 SHIPPED (AO 1638499) H3907AD 85EY67026 SHIPPED (AO 1638498) H3907AD 85EY67028 SHIPPED (AO 1638502) H3907AD 85EX67009 513732NM 5782013 85EX67002 419289D850 H3907AD 85EX67016 972969NM H3907AD 85EX67004 571472NM850 H3907AD 85EX67021 SHIPPED (GON 4640067) H3907AD 85EX67010 804289NM850A H3907AD 85EX67011 804281NM850A H3907AD 85EX67005 715717NMCT H3907AD 85EX67013 920456NM H3907AD 850X61004 030601NU06 H3907AD 850X61002 NS0001 H3907AD NT3BG1900001BH NS0401 5376204-50-1 85EX67003 PL1692NM05 H3907AD 85EX67014 PPR59076 H3907AD 850Y61007 SK1082NM02 H3907AD

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1262-2019
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for GE Healthcare, LLC? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for GE Healthcare, LLC: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about GE Healthcare, LLC →

Source: openFDA record for Z-1262-2019. Report a data issue: hello@regdatalab.com.