Z-1264-2019 Class II Terminated

Recalled by GE Healthcare, LLC — Waukesha, WI

FDA device recall Z-1264-2019 was initiated by GE Healthcare, LLC on Mar 20, 2019 and is designated Class II. Reason for recall: During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector… The recall status is terminated (terminated Oct 5, 2020). Affected quantity: 16 units.

Recall details

Product Type
Devices
Report Date
May 8, 2019
Initiation Date
Mar 20, 2019
Termination Date
Oct 5, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
16 units

Product description

NM/CT 870 CZT, Model/Catalogue/Code # H3906CW Product Usage: The GE Discovery NM/CT 670 CZT system is intended for general Nuclear Medicine imaging procedures for detection of radioisotope tracer uptake in the patient body. It includes a general purpose Nuclear Medicine system using a variety of scanning modes supported by various acquisition types, and a Computed Tomography system which is intended for enabling attenuation correction and anatomical localization of SPECT images and for standalone head, whole body, cardiac and vascular X-ray Computed Tomography applications.

Reason for recall

During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.

Distribution pattern

US and Austria, Canada, Croatia, France, Germany, India, Italy, Japan, Republic of Korea, Netherlands, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, UK

Code information

Mfg. Lot or Serial # System ID 87CY63017 SHIPPED (AO 1630621) 87CX63007 AH0101NM05 87CX63014 030078NU28 87CX63008 190020NU01 87CX63006 030240NU11 87CX63015 705743NM870 87CX63011 M4184879 87CX63003 M9886873 87CX63004 X171582003 87CX63009 M171582018 87CX63010 UC1796NU01 87CX63013 NO1068MN18 87CX63005 A004NU20 87CX63002 C028NU01 87CX63012 C028NU02 87CX63001 900010NU12

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1264-2019
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for GE Healthcare, LLC? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for GE Healthcare, LLC: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about GE Healthcare, LLC →

Source: openFDA record for Z-1264-2019. Report a data issue: hello@regdatalab.com.