Z-1265-2019 Class II Terminated
Recalled by GE Healthcare, LLC
— Waukesha, WI
FDA device recall Z-1265-2019 was initiated by GE Healthcare, LLC on Mar 20, 2019 and is designated Class II. Reason for recall: During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector… The recall status is terminated (terminated Oct 5, 2020). Affected quantity: 27 units.
Recall details
- Product Type
- Devices
- Report Date
- May 8, 2019
- Initiation Date
- Mar 20, 2019
- Termination Date
- Oct 5, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 27 units
Product description
NM/CT 870 DR Product Usage: The GE Discovery NM/CT 670 system is a medical tool intended for use by appropriately trained health professionals to aid in detecting, localizing, diagnosing of diseases and organ function for the evaluation of diseases, trauma, abnormalities, and disorders such as, but not limited to, cardiovascular disease, neurological disorders and cancer. The system output can also be used by the physician for staging and restaging of tumors, planning, guiding, and monitoring therapy. The GE Discovery NM/CT 670 system, combining Nuclear Medicine (NM) and Computed Tomography (CT) systems, is intended to produce
Reason for recall
During installation of certain Nuclear Medicine systems delivered with dismounted detectors, some detector screws may not have been secured correctly, leading to the possible loss of one detector mounting design redundancy.
Distribution pattern
US and Austria, Canada, Croatia, France, Germany, India, Italy, Japan, Republic of Korea, Netherlands, Poland, Portugal, Slovakia, South Africa, Spain, Sweden, Switzerland, UK
Code information
Mfg. Lot or Serial # System ID Model, Catalogue or Code # 870Y64037 SHIPPED (GON 4566933) H3100AS 870Y64040 SHIPPED (GON 4751375) H3100AS 870X64011 216844DR870 H3100AS 870X64024 802775D870 H3100AS 870X64002 206SCCANM H3100AS 870X64032 262741NUC H3100AS 870X64001 030020NU06 H3100AS 870Y64034 514345BD870 H3100AS 870Y64035 M54606132 H3100AS 870X64012 X634445001 H3100AS 870X64007 V2728113 H3100AS 870X64009 M4016255 H3100AS 870X64026 M40230109 H3100AS 870X64017 M40240121 H3100Q 870X64018 M40240122 H3100Q 870X64016 499221NU01 H3100AS 870X64022 A5173220 H3100AS NT3DG1800003BH NS0303 5376204-70-60-1 NT3DG1800002BH NS0302 5376204-70-60-1 870X64025 EMM0530 H3100AS 870Y64038 MST01300 H3100AS 870X64029 PL1031NM01 H3100AS 870X64027 PL0398NM01 H3100AS 870X64030 PL2190NM01 H3100AS 870X64028 PL0372NM01 H3100AS 870X64013 NP473631 H3100AS 870X64008 00669NUC03 H3100AS
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1265-2019
- openFDA file date
- Nov 27, 2026
- Loaded by RegDataLab
- Nov 28, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for GE Healthcare, LLC? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for GE Healthcare, LLC: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about GE Healthcare, LLC →
Use it in ChatGPT
Get a free API key
How the connector works