Z-1270-2018 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- April 4, 2018
- Initiation Date
- December 18, 2017
- Termination Date
- April 9, 2019
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
Product Description
cobas e 411 Immunoassay Analyzer (Discrete Photometric Analyzer for Clinical use), Catalog Numbers: 04775201001 (rack system), 04775279001 (disk system), 04775201973 (rack system) & 04775279973 (disk system).
Reason for Recall
Complaints have been received concerning questionable results on analyzers due to customers utilizing 13 mm sample tubes without tube adapters.
Distribution Pattern
US Nationwide
Code Information
None