Z-1281-2026 Class II Ongoing

Recalled by Olympus Corporation Of The Americas — Center Valley, PA

FDA device recall Z-1281-2026 was initiated by Olympus Corporation Of The Americas on Jan 7, 2026 and is designated Class II. Reason for recall: Devices which did not undergo thermoforming could deform and lose performance. The recall status is ongoing. Affected quantity: 5089 units (30 US, 5059 OUS).

Recall details

Product Type
Devices
Report Date
Feb 11, 2026
Initiation Date
Jan 7, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5089 units (30 US, 5059 OUS)

Product description

Product Name: Single Use 3-Lumen Sphincterotome V (Distal Wire guided) Model/Catalog Number: KD-VC411Q-0330; Software Version: N/A Product Description: These instruments have been designed to be used with an Olympus endoscope and guidewire for papillotomy using high-frequency current.

Reason for recall

Devices which did not undergo thermoforming could deform and lose performance.

Distribution pattern

Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, DC, DE, FL, GA, GU, HI, IA, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and International: TBD.

Code information

Model/Catalog Number: KD-VC411Q-0330; Material REF: (1) N5391130, (2) N5391110, (3) N5777430; UDI-DI: (1) 04953170371097, (2) 04953170371097, (3) 04953170371097; All Lots with a valid expiration date

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1281-2026
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1281-2026. Report a data issue: hello@regdatalab.com.