Z-1288-2015 Class II Terminated

Recalled by Ossur H / F — Reykjavik

FDA device recall Z-1288-2015 was initiated by Ossur H / F on Mar 4, 2015 and is designated Class II. Reason for recall: The threads of the outer tube of the shock module of LP Rotate are too short, resulting in insecure assembly. Use of the product may cause the patient to fall with continued use, as the shock module… The recall status is terminated (terminated Oct 28, 2015). Affected quantity: 2,194 (1,464 in U.S. and 730 internationally).

Recall details

Product Type
Devices
Report Date
Mar 25, 2015
Initiation Date
Mar 4, 2015
Termination Date
Oct 28, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2,194 (1,464 in U.S. and 730 internationally)

Product description

LP Rotate Foot System prosthetic device for fitting lower extremity amputations. -Low profile foot module with adjustable shock absorber and rotation with progressive stiffening -Build height: with male pyramid adapter 145 - 153mm (5 11/16 - 6 ) with female pyramid adapter 153 165mm (6 - 6 1/2) -Male pyramid adapter -Female pyramid adapter -Maintenance free, no lubrication required -Maximum weight: 147kg (325lbs) -Available in sizes 22-30

Reason for recall

The threads of the outer tube of the shock module of LP Rotate are too short, resulting in insecure assembly. Use of the product may cause the patient to fall with continued use, as the shock module can become loose from the foot blade.

Distribution pattern

Nationwide within the U.S. and internationally to the following foreign countries; Belgium, United Kingdom, Germany, Czech Republic, France, Israel, Poland, Spain, Australia, Japan, Mexico, Bulgaria, Canada, South Korea, Sweden, Finland,

Code information

LP Rotate, sizes 25-27 cat.1-6. Product numbers LRP062, LRF062, LRPU62, LRFU62. All from manufacturing lot number HF141022

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1288-2015
openFDA file date
Nov 17, 2026
Loaded by RegDataLab
Nov 18, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1288-2015. Report a data issue: hello@regdatalab.com.