Z-1288-2015 Class II Terminated
FDA device recall Z-1288-2015 was initiated by Ossur H / F on Mar 4, 2015 and is designated Class II. Reason for recall: The threads of the outer tube of the shock module of LP Rotate are too short, resulting in insecure assembly. Use of the product may cause the patient to fall with continued use, as the shock module… The recall status is terminated (terminated Oct 28, 2015). Affected quantity: 2,194 (1,464 in U.S. and 730 internationally).
Recall details
- Product Type
- Devices
- Report Date
- Mar 25, 2015
- Initiation Date
- Mar 4, 2015
- Termination Date
- Oct 28, 2015
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 2,194 (1,464 in U.S. and 730 internationally)
Product description
Reason for recall
Distribution pattern
Nationwide within the U.S. and internationally to the following foreign countries; Belgium, United Kingdom, Germany, Czech Republic, France, Israel, Poland, Spain, Australia, Japan, Mexico, Bulgaria, Canada, South Korea, Sweden, Finland,
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1288-2015
- openFDA file date
- Nov 17, 2026