Z-1292-2018 Class II Terminated

Recalled by GE Healthcare, LLC — Waukesha, WI

Recall Details

Product Type
Devices
Report Date
April 11, 2018
Initiation Date
December 15, 2017
Termination Date
April 21, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1742

Product Description

Uterine Electromyographic Monitor - Monica Novii Wireless Patch System Product Usage: The Novii Interface is an accessory to the Novii Pod which provides a means of interfacing the wireless output of the Novii Pod to the transducer inputs of a CTG Fetal monitor via three cables. The Novii Interface enables signals collected by the Novii Pod to be printed and displayed on a CTG Fetal Monitor and sent on to a central network, if connected.

Reason for Recall

if the cables of the Monica Novii System are removed by a user, and then incorrectly re-connected the Fetal Heart Rate (FHR) and Maternal Heart Rate (MHR) signals would be displayed incorrectly (swapped).

Distribution Pattern

Worldwide - US Nationwide Distribution.

Code Information

Model # - 107-PT-020 ( all units distributed from January 2016- November 2017)