Z-1293-2018 Class II Terminated

Recalled by GE Healthcare, LLC — Waukesha, WI

FDA device recall Z-1293-2018 was initiated by GE Healthcare, LLC on Dec 15, 2017 and is designated Class II. Reason for recall: if the cables of the Monica IF24 System are removed by a user, and then incorrectly re-connected the Fetal Heart Rate (FHR) and Maternal Heart Rate (MHR) signals would be displayed incorrectly… The recall status is terminated (terminated Apr 21, 2020). Affected quantity: 355.

Recall details

Product Type
Devices
Report Date
Apr 11, 2018
Initiation Date
Dec 15, 2017
Termination Date
Apr 21, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
355

Product description

Uterine Electromyographic Monitor -Monica IF24 Interface System Product Usage: The Monica IF24 CTG Interface Device is an accessory to the Monica AN24 which provides a means of interfacing the wireless output of the Monica AN24 to the transducer inputs of a CTG Fetal monitor. The Monica IF24 CTG Interface Device enables signals collected by the Monica AN24 to be printed and displayed on a CTG Fetal Monitor and sent on to a central network, if connected. The device is intended for use by healthcare professionals in a clinical setting.

Reason for recall

if the cables of the Monica IF24 System are removed by a user, and then incorrectly re-connected the Fetal Heart Rate (FHR) and Maternal Heart Rate (MHR) signals would be displayed incorrectly (swapped).

Distribution pattern

Worldwide - US Nationwide Distribution.

Code information

Model # - 105-PT-001 ( all units distributed from October 2011 - May 2016)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1293-2018
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for GE Healthcare, LLC? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1293-2018. Report a data issue: hello@regdatalab.com.