Recalled by Bio-Rad Laboratories, Inc. — Hercules, CA

FDA device recall Z-1299-2014 was initiated by Bio-Rad Laboratories, Inc. on Feb 27, 2014 and is designated Class II. Reason for recall: On a rare occasion the D-10 software might generate an extra result by duplicating the result form the previous sample and the Sample ID/Injection number from the next sample. The recall status is terminated (terminated Jun 4, 2014). Affected quantity: 483 instruments in US. OUS to be determined.

Recall details

Product Type
Devices
Report Date
Apr 2, 2014
Initiation Date
Feb 27, 2014
Termination Date
Jun 4, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
483 instruments in US. OUS to be determined.

Product description

D-10" Hemoglobin Testing System, Model number: 220-0220 Manufactured by: Bio-Rad Laboratories, Inc. 4000 Alfred Nobel Drive Hercules, CA 94545 The D-10 Hemoglobin Testing System is a single module designed for hemoglobin analysis in mid-volume laboratories. The D-10 is a fully integrated system for sample dilution, processing and analysis of hemoglobin, designed for use with specific Bio-Rad reagent kits. The D-10 incorporates the use of a dedicated software system for instrument control, data collection and analysis.

Reason for recall

On a rare occasion the D-10 software might generate an extra result by duplicating the result form the previous sample and the Sample ID/Injection number from the next sample.

Distribution pattern

Worldwide Distribution - USA (nationwide) and Internationally to Canada, Japan, China, India, SriLanka, Argentina, Paraguay, Uruguay, Chile, Peru ,Bolivia ,Mexico ,Panama, ,Costa Rica ,Ecuador, Venezuela, Colombia, Brazil, Jamaica, Austria, Switzerland, France, Germany, Greece, Portugal, Spain, Italy, UK, Belgium, Sweden Norway, Finland ,Israel/Palestine, Tunisia, Algeria, Egypt, ,Morocco, Jordan, Yemen, Iraq, Kuwait, Qatar, Oman, Lybia, Lebanon ,UAE, Saudi Arabia, Pakistan, Afghanistan, Turkey, South Africa, Chad, Kenya, Benin, Burkina Faso, Cameroon, Gabon, Senegal, Ghana, Mali, Ivory Coast, Nigeria, Djibouti, Czech Republic, Slovakia, Slovenia, Serbia, Hungary, Romania, Poland, Russia, Ukraine, Kazakhstan, Bulgaria, Belarus, Georgia, Mauritius, Maldives, Vietnam, Philippines, Korea, Thailand, Hong Kong , Indonesia, Taiwan, Malaysia, Singapore, New Zealand, and Australia.

Code information

Model number: 220-0220

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1299-2014
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Bio-Rad Laboratories, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Bio-Rad Laboratories, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Bio-Rad Laboratories, Inc. →

Source: openFDA record for Z-1299-2014. Report a data issue: hello@regdatalab.com.