Z-1301-2025 Class II Ongoing

Recalled by GE Medical Systems, LLC — Waukesha, WI

Recall Details

Product Type
Devices
Report Date
March 19, 2025
Initiation Date
February 18, 2025
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1 unit

Product Description

GE Healthcare Revolution Apex Select, System, X-ray, Tomography, Computed

Reason for Recall

GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.

Distribution Pattern

Worldwide distribution.

Code Information

System ID Number 907729APEX1, UDI (01)00195278460271(11)231100(21)REV2X2300082CN, Mfg. Lot/Serial Number REV2X2300082CN