Z-1301-2025 Class II Ongoing

Recalled by GE Medical Systems, LLC — Waukesha, WI

FDA device recall Z-1301-2025 was initiated by GE Medical Systems, LLC on Feb 18, 2025 and is designated Class II. Reason for recall: GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant… The recall status is ongoing. Affected quantity: 1 unit.

Recall details

Product Type
Devices
Report Date
Mar 19, 2025
Initiation Date
Feb 18, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1 unit

Product description

GE Healthcare Revolution Apex Select, System, X-ray, Tomography, Computed

Reason for recall

GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.

Distribution pattern

Worldwide distribution.

Code information

System ID Number 907729APEX1, UDI (01)00195278460271(11)231100(21)REV2X2300082CN, Mfg. Lot/Serial Number REV2X2300082CN

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1301-2025
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1301-2025. Report a data issue: hello@regdatalab.com.