Z-1308-2021 Class II Terminated

Recalled by Biomerieux Inc — Hazelwood, MO

FDA device recall Z-1308-2021 was initiated by Biomerieux Inc on Jan 28, 2021 and is designated Class II. Reason for recall: bioM¿rieux has identified a potential safety risk worst case of a false susceptible erroneous test result associated with this event. The problem is when HL7 Connection is used, the results in the… The recall status is terminated (terminated Jan 12, 2023). Affected quantity: 6777 systems.

Recall details

Product Type
Devices
Report Date
Mar 31, 2021
Initiation Date
Jan 28, 2021
Termination Date
Jan 12, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6777 systems

Product description

VITEK 2 Compact (version numbers 8.01, 8.02 (8.0x) and 9.01, 9.02 (9.0x) with the external communication configured to HL7 (system, test, automated, antimicrobial susceptibility, short incubation). VITEK 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The VITEK 2 utilizes growth-based biochemical patterns to determine identification. The VITEK 2 provides minimal inhibitory concentration (MIC) results for most organism/drug combinations as well as a category interpretation (S, I, or R) consistent with accepted criteria.

Reason for recall

bioM¿rieux has identified a potential safety risk worst case of a false susceptible erroneous test result associated with this event. The problem is when HL7 Connection is used, the results in the VITEK¿ 2 do not match the results sent to the LIS and the Laboratory Technician would need to change results in the LIS to match those in the VITEK¿ 2. If the isolate is sent more than once to LIS, a software defect prevents the system from sending the expertised interpretation results and the Therapeutic Corrections (TC) is not sent to Laboratory Information Systems (LIS).

Distribution pattern

Worldwide distribution - No US consignees affected and the countries of Andorra, Austria, Bangladesh, Bermuda, Brazil, Brunei Darussalam, El Salvador, French Guiana, French Polynesia, Gibraltar, Greece, Guadeloupe, Guam, Holy See (Vatican City State), Hong Kong, Hungary, India, Indonesia, Ireland, Lao People's Democratic Republic, Luxembourg, Macao, Malaysia, Martinique, Mexico, Myanmar, New Caledonia, New Zealand, Nigeria, Philippines, Reunion, San Marino, Slovakia, Sri Lanka, Sweden, Switzerland, Thailand, Tunisia, Uruguay.

Code information

VITEK 2 and VITEK 2 Compact (version numbers 8.01, 8.02 (8.0x) and 9.01, 9.02 (9.0x) with the external communication configured to HL7. Each instrument has unique serial number. (System: VITEK 2 Compact 60, Reference Number: 27560, Catalog Number: C57, UDI: 03573026136796); (System: VITEK 2 Compact 30, Reference Number: 27560R, Catalog Number: C57, UDI: 03573026147501); (System: VITEK 2 Compact 30, Reference Number: 27530, Catalog Number: C57, UDI: 03573026136789); (System: VITEK 2 Compact 30, Reference Number: 27530R, Catalog Number: C57, UDI: 03573026147495); (System: VITEK 2 Compact 15, Reference Number: 27415, Catalog Number: C57, UDI: 03573026220624); (System: VITEK 2 Compact 15, Reference Number: 27415R, Catalog Number: C57, UDI: 03573026220631)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1308-2021
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Biomerieux Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Biomerieux Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Biomerieux Inc →

Source: openFDA record for Z-1308-2021. Report a data issue: hello@regdatalab.com.