Z-1364-2022 Class II Terminated

Recalled by Fresenius Medical Care Holdings, Inc. — Waltham, MA

FDA device recall Z-1364-2022 was initiated by Fresenius Medical Care Holdings, Inc. on Aug 3, 2021 and is designated Class II. Reason for recall: AquaBplus HF module on the AquaBplus 2000 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an… The recall status is terminated (terminated Apr 10, 2024). Affected quantity: 2 units.

Recall details

Product Type
Devices
Report Date
Jul 20, 2022
Initiation Date
Aug 3, 2021
Termination Date
Apr 10, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2 units

Product description

AquaBplus Water Purification System, Part Number 24-2000-0, AquaBplus 2500 - Typ USA (G02040108-US), AquaBplus 2000 - Typ USA (G02040114-US), AquaBplus HF -Typ USA (G02040120FMC-US)

Reason for recall

AquaBplus HF module on the AquaBplus 2000 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2000 device. The 3-ring main and fluid fly loops the heat disinfection may not be as effective as a 2-ring main installation

Distribution pattern

US Nationwide distribution in the states of NC and WV.

Code information

UDI-DI: (01)04251285200227, (01)04251285200289, (01)0425128520; Serial Numbers: 7BPS1621 7BDS0783 7BFS0517 7BPS1611 7BDS0776 7BFS0502

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1364-2022
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1364-2022. Report a data issue: hello@regdatalab.com.