Z-1366-2022 Class II Ongoing

Recalled by Wright Medical Technology, Inc. — Memphis, TN

FDA device recall Z-1366-2022 was initiated by Wright Medical Technology, Inc. on Jun 14, 2022 and is designated Class II. Reason for recall: The plate cutter is unable to meet sterilization requirements when it is sterilized in a double-wrapped configuration according to the Option 1 Steam Sterilization parameters detailed in instructions… The recall status is ongoing. Affected quantity: 231.

Recall Details

Product Type
Devices
Report Date
Jul 20, 2022
Initiation Date
Jun 14, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
231

Product Description

Wright EVOLVE TRIAD Plate Cutter, REF 49510120

Reason for Recall

The plate cutter is unable to meet sterilization requirements when it is sterilized in a double-wrapped configuration according to the Option 1 Steam Sterilization parameters detailed in instructions for use

Distribution Pattern

Nationwide and PR, and Australia, Canada, Chile, Colombia, Hong Kong, Puerto Rico

Code Information

Lot codes: 955401, 1168853, 1349302, 1956936, 2486857

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1366-2022
openFDA file date
Nov 3, 2026
Loaded by RegDataLab
Nov 4, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1366-2022. Report a data issue: hello@regdatalab.com.