Z-1402-2013 Class II Terminated

Recalled by Elekta, Inc. — Atlanta, GA

FDA device recall Z-1402-2013 was initiated by Elekta, Inc. on Apr 30, 2013 and is designated Class II. Reason for recall: Unexpected movement during beam delivery may not be detected if there are incorrect static tolerances for Linac Geometric Axes loaded in Integrity R3.0. The recall status is terminated (terminated Jul 26, 2017). Affected quantity: 31.

Recall details

Product Type
Devices
Report Date
Jun 5, 2013
Initiation Date
Apr 30, 2013
Termination Date
Jul 26, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
31

Product description

Integrity 3.0 Product Usage: The Agility multileaf collimator is indicated for use when additional flexibility is required in conforming the radiation beam to the anatomy to be exposed.

Reason for recall

Unexpected movement during beam delivery may not be detected if there are incorrect static tolerances for Linac Geometric Axes loaded in Integrity R3.0.

Distribution pattern

USA Nationwide including the states of CA, FL, IL, MD, MA, NJ, NY, NC, ND, OH, OR, PA, TN, TX, VA, WA, DC and Puerto Rico

Code information

325134, 325254, 325082, 325159, 325268, 325237, 325331, 325267, 325236, 325314, 325127, 325261, 325286, 325077, 325269, 325129, 325296, 325273, 325139, 325201, 325137, 325184, 325161, 325076, 325086, 325112, 325081, 325083, 325099, 325110, 325271

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1402-2013
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1402-2013. Report a data issue: hello@regdatalab.com.