Z-1402-2013 Class II Terminated
FDA device recall Z-1402-2013 was initiated by Elekta, Inc. on Apr 30, 2013 and is designated Class II. Reason for recall: Unexpected movement during beam delivery may not be detected if there are incorrect static tolerances for Linac Geometric Axes loaded in Integrity R3.0. The recall status is terminated (terminated Jul 26, 2017). Affected quantity: 31.
Recall details
- Product Type
- Devices
- Report Date
- Jun 5, 2013
- Initiation Date
- Apr 30, 2013
- Termination Date
- Jul 26, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 31
Product description
Integrity 3.0 Product Usage: The Agility multileaf collimator is indicated for use when additional flexibility is required in conforming the radiation beam to the anatomy to be exposed.
Reason for recall
Unexpected movement during beam delivery may not be detected if there are incorrect static tolerances for Linac Geometric Axes loaded in Integrity R3.0.
Distribution pattern
USA Nationwide including the states of CA, FL, IL, MD, MA, NJ, NY, NC, ND, OH, OR, PA, TN, TX, VA, WA, DC and Puerto Rico
Code information
325134, 325254, 325082, 325159, 325268, 325237, 325331, 325267, 325236, 325314, 325127, 325261, 325286, 325077, 325269, 325129, 325296, 325273, 325139, 325201, 325137, 325184, 325161, 325076, 325086, 325112, 325081, 325083, 325099, 325110, 325271