Z-1413-2017 Class II Terminated

Recalled by Encore Medical, Lp — Austin, TX

FDA device recall Z-1413-2017 was initiated by Encore Medical, Lp on Feb 9, 2017 and is designated Class II. Reason for recall: During the Turon assembly, the impaction forces caused the polymer, black acetal copolymer from the Impaction Fixture to wear off on the lateral surface of the Humeral Stem titanium plasma spray… The recall status is terminated (terminated Oct 20, 2017). Affected quantity: 626 units.

Recall details

Product Type
Devices
Report Date
Mar 15, 2017
Initiation Date
Feb 9, 2017
Termination Date
Oct 20, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
626 units

Product description

RSP Impaction Fixture

Reason for recall

During the Turon assembly, the impaction forces caused the polymer, black acetal copolymer from the Impaction Fixture to wear off on the lateral surface of the Humeral Stem titanium plasma spray coating.

Distribution pattern

US, South Korea, Australia, Canada, United Kingdom/Ireland, Germany

Code information

109931L01, 109931L02, 115670L15, 115670L16, 128092L08, 128092L09, 137917L16, 167829L06, 52258L01, 52258L01A, 67428L01, 76386L01, 76386L02, 81722L01, 81722L02

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1413-2017
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1413-2017. Report a data issue: hello@regdatalab.com.