Z-1413-2017 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- March 15, 2017
- Initiation Date
- February 9, 2017
- Termination Date
- October 20, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 626 units
Product Description
RSP Impaction Fixture
Reason for Recall
During the Turon assembly, the impaction forces caused the polymer, black acetal copolymer from the Impaction Fixture to wear off on the lateral surface of the Humeral Stem titanium plasma spray coating.
Distribution Pattern
US, South Korea, Australia, Canada, United Kingdom/Ireland, Germany
Code Information