Z-1424-2022 Class II Terminated

Recalled by Randox Laboratories Ltd. — Crumlin (North), N/A

FDA device recall Z-1424-2022 was initiated by Randox Laboratories Ltd. on Jun 22, 2022 and is designated Class II. Reason for recall: Product fails to meet the performance claims quoted on the kit insert: Manual Procedure when calibrating using the standard provided in the kit, the change in absorbance is lower than expected… The recall status is terminated (terminated May 10, 2024). Affected quantity: 15 kits.

Recall Details

Product Type
Devices
Report Date
Jul 27, 2022
Initiation Date
Jun 22, 2022
Termination Date
May 10, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
15 kits

Product Description

Randox Cholesterol- For the quantitative in vitro determination of Cholesterol in serum and plasma. Catalog Number: CH2O0

Reason for Recall

Product fails to meet the performance claims quoted on the kit insert: Manual Procedure when calibrating using the standard provided in the kit, the change in absorbance is lower than expected generating internal quality control results high outside range; Cholesterol (CHOL) Instrument Specific Application using the recommended calibration material CAL2351, the product fails to meet its linearity performance claim quoted on the kit inserts. The linearity is reduced by up to -35% however Quality Control results will still fall within assigned ranges. This may lead to delayed results.

Distribution Pattern

Nationwide Distribution - MD, OR, PR, MT, WV

Code Information

GTIN : 05055273201130 Lot Number: 586177 Expiry Date: 28th Jan 2024

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1424-2022
openFDA file date
Oct 23, 2026
Loaded by RegDataLab
Oct 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.