Z-1424-2022 Class II Terminated
FDA device recall Z-1424-2022 was initiated by Randox Laboratories Ltd. on Jun 22, 2022 and is designated Class II. Reason for recall: Product fails to meet the performance claims quoted on the kit insert: Manual Procedure when calibrating using the standard provided in the kit, the change in absorbance is lower than expected… The recall status is terminated (terminated May 10, 2024). Affected quantity: 15 kits.
Recall Details
- Product Type
- Devices
- Report Date
- Jul 27, 2022
- Initiation Date
- Jun 22, 2022
- Termination Date
- May 10, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 15 kits
Product Description
Randox Cholesterol- For the quantitative in vitro determination of Cholesterol in serum and plasma. Catalog Number: CH2O0
Reason for Recall
Product fails to meet the performance claims quoted on the kit insert: Manual Procedure when calibrating using the standard provided in the kit, the change in absorbance is lower than expected generating internal quality control results high outside range; Cholesterol (CHOL) Instrument Specific Application using the recommended calibration material CAL2351, the product fails to meet its linearity performance claim quoted on the kit inserts. The linearity is reduced by up to -35% however Quality Control results will still fall within assigned ranges. This may lead to delayed results.
Distribution Pattern
Nationwide Distribution - MD, OR, PR, MT, WV
Code Information
GTIN : 05055273201130 Lot Number: 586177 Expiry Date: 28th Jan 2024