Z-1427-2023 Class I Ongoing
FDA device recall Z-1427-2023 was initiated by Cook Incorporated on April 7, 2023 and is designated Class I. Reason for recall: There were reports from customers that the tracheostomy in the sets and trays is not making a secure connection with the 15mm cap and other 15mm circuit components and accessories. This resulted in a… The recall status is ongoing. Affected quantity: 3.053 units.
Recall Details
- Product Type
- Devices
- Report Date
- May 17, 2023
- Initiation Date
- April 7, 2023
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3.053 units
Product Description
Blue Rhino G2-Multi Percutaneous Tracheostomy Introducer Set, Reference Part Numbers C-PTIS-100-HC-G-EU-FLEX7.5 (G57695), C-PTIS-100-HC-G-EU-FLEX8.5 (G57696), C-PTIS-100-HC-G-NA-FLEX7.5 (G57691), C-PTIS-100-HC-G-NA-FLEX8.5 (G57692)
Reason for Recall
There were reports from customers that the tracheostomy in the sets and trays is not making a secure connection with the 15mm cap and other 15mm circuit components and accessories. This resulted in an unsecure connection between the device connector and circuit components, cap, or accessories.
Distribution Pattern
Worldwide distribution - US Nationwide.
Code Information
C-PTIS-100-HC-G-EU-FLEX7.5 (G57695): UDI-DI 00827002576957, Lot Number 14842940; C-PTIS-100-HC-G-EU-FLEX8.5 (G57696): UDI-DI 00827002576964, Lot Numbers 14727192, 14332964, 14335896, 14380758, 14380755, 14332978, 14332973, 14335892, 14335893, 14464000, 14332985, 14464008, 14463988, 14503228, 14503224, 14503222, 14503229, 14503227, 14742869, 14742870, 14727207, 14742873; C-PTIS-100-HC-G-NA-FLEX7.5 (G57691): UDI-DI 00827002576919, Lot Numbers 15063922, 15063923, 15103482, 14227740, 14221180, 14204927, 14227743, 14221125, 14221198, 14221190, 14163328, 14250265, 14250269, 14317245, 14317213, 14345634, 14376893, 14633088, 14633089, 14794549, 14794548, 14794547, 14874486, 14874484, 14926500, 14967737, 14926505, 14967739, 14967741, 14221180; C-PTIS-100-HC-G-NA-FLEX8.5 (G57692): UDI-DI 00827002576926, Lot Numbers 14608362, 14750607, 14894816, 14264399, 14299799, 14299802, 14299803, 14345667, 14365693, 14365699, 14365696, 14365694, 14365701, 14345687, 14345587, 14456438, 14456443, 14456431, 14407846, 14750610, 14488394, 14488418, 14608361, 14608363, 14750618, 14608360, 14961742, 14961716, 15089037, 15089038, 15136113