Z-1438-2021 Class II Ongoing

Recalled by Heartware, Inc. — Miami Lakes, FL

FDA device recall Z-1438-2021 was initiated by Heartware, Inc. on February 26, 2021 and is designated Class II. Reason for recall: IFU and Patient Manual updated to describe HVAD Controller Connection Verification. The labeling is being corrected to reflect the actual verified behavior of a successful connector to the controller,… The recall status is ongoing. Affected quantity: 23123 (11292 US).

Recall Details

Product Type
Devices
Report Date
April 28, 2021
Initiation Date
February 26, 2021
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
23123 (11292 US)

Product Description

Instructions for Use and Patient Manual for HeartWare HVAD System. Accompanies these devices: Heartware HVAD pump implant kits Models 1100, 1101, 1102, 1103, 1104, 1104JP, 1205, and MCS1705PU. Heartware HVAD controller kits Models 1403US, 1407AU, 1407CA, 1407CH, 1407DE, 1407GB, 1407IL, 1407IN, 1407IT, 1407JP, 1407KR, 1420, and 1420JP. Heartware HVAD controller AC adapter Models 1430AR, 1430AU, 1430CA, 1430CH, 1430DE, 1430GB, 1430IL, 1430IN, 1430IT, 1430JP, and 1430US. Heartware HVAD controller DC Adapter Model 1440. Heartware HVAD data cable - monitor Model 1575. Heartware HVAD battery pack Models 1650, 1650CA-CLIN, and 1650DE. Heartware HVAD patient pack Model 1475. Heartware HVAD waist pack Models 2050, 2050IL, and 2050US. Heartware HVAD shoulder pack Models 2060, 2060IL, and 2060US.

Reason for Recall

IFU and Patient Manual updated to describe HVAD Controller Connection Verification. The labeling is being corrected to reflect the actual verified behavior of a successful connector to the controller, which is a "beep" and not "audible click", to clarify under what alarm conditions a beep will not be heard, and identify alternate confirmation methods (visual indicator) for successful connection. NOTE: this update is included in the February 2021 Urgent Medical Device Notice also covered in recalls 87456 and 87603

Distribution Pattern

Worldwide distribution - US Nationwide distribution and the countries of Italy, Great Britain, Germany, Japan.

Code Information

Model, GTIN: 1100, No GTIN Available; 1101, No GTIN Available; 1102, No GTIN Available; 1103, 00888707000017; 1104, No GTIN Available; 1104JP, 00888707000697; 1205, No GTIN Available; MCS1705PU, 00888707005364; 1403US, 00888707000475; 1407AU, 00888707001663; 1407CA, 00888707002851; 1407CH, 00888707001694; 1407DE, 00888707001700; 1407GB, 00888707001717; 1407IL, 00888707001724; 1407IN, 00888707001731; 1407IT, 00888707001748; 1407JP, 00888707001755; 1407KR, 00888707001762; 1420, 00888707000420; 1420JP, 00888707000437; 1430AR, 00888707000826; 1430AU, 00888707000833; 1430CA, 00888707000857; 1430CH, 00888707000871; 1430DE, 00888707000864; 1430GB, 00888707000888; 1430IL, 00888707000901; 1430IN, 00888707000505; 1430IT, 00888707000895; 1430JP, 00888707000918; 1430US, 00888707000307; 1440, 00888707001885; 1575, 00888707008051; 1650, 00888707000376; 1650CA-CLIN, 00888707001588; 1650DE, 00888707001373; 1475, 00888707008037; 2050, 00888707008082; 2050IL, 00888707001410; 2050US, 00888707000048; 2060, 00888707008099; 2060IL, 00888707001427; 2060US, 00888707000055;