Z-1440-2013 Class II Terminated

Recalled by Carl Zeiss Meditec, Inc. — Dublin, CA

FDA device recall Z-1440-2013 was initiated by Carl Zeiss Meditec, Inc. on Apr 18, 2013 and is designated Class II. Reason for recall: Carl Zeiss Meditec is conducting a field corrective action for its OPMI Pentero C surgical microscope system. During a routine inspection, one microscope installation was found to have 3 (three)… The recall status is terminated (terminated Jun 26, 2013). Affected quantity: 6 units.

Recall details

Product Type
Devices
Report Date
Jun 12, 2013
Initiation Date
Apr 18, 2013
Termination Date
Jun 26, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6 units

Product description

Zeiss brand OPMI Pentero C, Surgical microscope and accessories, Model: OPMI Pentero C, Product Numbers: 302583-9001-000 (PENTERO C), and 305961-0000-00 (Ceiling Mount for PENTERO C), Product Usage: A surgical microscope and accessories is an AC-powered device intended for use during surgery to provide a magnified view of the surgical field.

Reason for recall

Carl Zeiss Meditec is conducting a field corrective action for its OPMI Pentero C surgical microscope system. During a routine inspection, one microscope installation was found to have 3 (three) broken screws out of 6 (six) in the lower section of the ceiling mount.

Distribution pattern

US Nationwide Distribution.

Code information

Serial Numbers: 6725000708, 6725000709, 6725000710, 6725000713, 6725100725, 6725120813

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1440-2013
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Carl Zeiss Meditec, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-1440-2013. Report a data issue: hello@regdatalab.com.