Z-1452-2013 Class I Terminated

Recalled by Symbios Medical Products, LLC — Indianapolis, IN

Recall Details

Product Type
Devices
Report Date
June 12, 2013
Initiation Date
May 10, 2013
Termination Date
March 7, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
33, 543 in all sizes/ kits

Product Description

Part 510031 Disposable Single Infusion Pump, quantity Includes... TOTAL VOLUME: 150mL, 2 mL/hr continuous Delivery Time : App. 72 hrs when filled to capacity Sterile EO... found in the following GoPump Kits: GoPump Kit 510042, w/ Epidural Catheter GoPump Kit 510042 - BP, w/ Epidural Catheter, BIOPATCH GoPump Kit 510110, w/2.5" Fenestrated Catheter GoPumpKit 510110-BP, w/2.5" Fenestrated Catheter,BioPatch GoPumpKit 510350 w/ Epidural Catheter GoPump Kit 510350-BP w/ Epidural Catheter, BIOPATCH GoPumpKit 510556-BP , 2.5 catheter, BIOPATCH Symbios GOPump Rapid Recovery System is a local pain mangagement system that consits of a small balloon that is inflated with a local anesthetic medication.

Reason for Recall

Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being expanded after receiving additional reports relating to excessively high flow rates on the 150 mL reservoir, 2 mL/hr GOPump (PN510031), and the

Distribution Pattern

US Nationwide distribution.

Code Information

Pump 510031, Lot 11-l 00106 to Lot 12-101133 Kit 510042, lots 11-100251 to 13-100128 Kit 510042-BP , lot 12-101300 Kit 510110, lot 11-100664 Kit 510110-BP, Lots 11-100838 to 12-101560 Kit 510350, Lots 11-100353 to 12-100233 Kit 510350-BP, Lots12-100429 to 13-100047 Kit 510556-BP Lots 11-100779 to 12-101552