Z-1459-2013 Class I Terminated

Recalled by Symbios Medical Products, LLC — Indianapolis, IN

Recall Details

Product Type
Devices
Report Date
June 12, 2013
Initiation Date
May 10, 2013
Termination Date
March 7, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
33, 543 for all kits

Product Description

Part 510490 Disposable Infusion Pump, ...Elastomeric Pump 600 mL( 300 ml per side)( 6mL/hr ),... found in the following kits GoBlock Kit 510491, Disposable Single Infusion Pump for drug delivery

Reason for Recall

Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being expanded after receiving additional reports relating to excessively high flow rates on the 150 mL reservoir, 2 mL/hr GOPump (PN510031), and the

Distribution Pattern

US Nationwide distribution.

Code Information

Pump Part 510490, Lot 11-100379- Lot 12-101132 found in: GoBlock Kit 510491, Lot 11-100601 to 13-100705