Z-1454-2013 Class I Terminated

Recalled by Symbios Medical Products, LLC — Indianapolis, IN

Recall Details

Product Type
Devices
Report Date
June 12, 2013
Initiation Date
May 10, 2013
Termination Date
March 7, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
all kits 33, 543

Product Description

Part 510180 Disposable Single Infusion Pump, ... Elastomeric Pump 300 mL, 2mL/hr,... found in : GoPump Kit 510201-BP, 5 fenestrated catheter & BIOPATCH; Go Pump Kit 510349-BP, Epidural Catheter, BioPatch; Go Pump Kit 510642 Disposable Single Infusion Pump, Symbios GOPump Rapid Recovery System is a local pain mangagement system that consists of a small balloon that is inflated with a local anesthetic medication. The medication is delivered via tubing to the surgical site at a very slow rate

Reason for Recall

Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being expanded after receiving additional reports relating to excessively high flow rates on the 150 mL reservoir, 2 mL/hr GOPump (PN510031), and the

Distribution Pattern

US Nationwide distribution.

Code Information

Pump Part 510180, Lot 11-100453 to 12-101003 found in: Go Pump Kit 510201-BP. Lot 12-100108 to 12-101538 Go Pump Kit 510349-BP. Lot 12-100935 Go Pump Kit 510642, Lot 12-100616 to 13-100590