Z-1457-2013 Class I Terminated

Recalled by Symbios Medical Products, LLC — Indianapolis, IN

Recall Details

Product Type
Devices
Report Date
June 12, 2013
Initiation Date
May 10, 2013
Termination Date
March 7, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
N/A

Product Description

Part 510449 Disposable Infusion Pump, ...Elastomeric Pump 600 mL( 8mL/hr ),... found in the following kits GoBlock Kit 510472, Disposable Single Infusion Pump. Symbios GOPump Rapid Recovery System is a local pain mangagement system that consists of a small balloon that is inflated with a local anesthetic medication.

Reason for Recall

Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being expanded after receiving additional reports relating to excessively high flow rates on the 150 mL reservoir, 2 mL/hr GOPump (PN510031), and the

Distribution Pattern

US Nationwide distribution.

Code Information

Pump Part 510449, Lot 11-100271 - Lot 130100137 found in: GoBlock Kit 510472, Lot 11-100385 to 13-100354