Z-1453-2013 Class I Terminated

Recalled by Symbios Medical Products, LLC — Indianapolis, IN

Recall Details

Product Type
Devices
Report Date
June 12, 2013
Initiation Date
May 10, 2013
Termination Date
March 7, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
all kits 33,543

Product Description

Part 510076 Disposable Dual Infusion Pump, ...Elastomeric Dual Pump 300 mL(150 ml/side) 2mL/hr/side,... Sterile, found in the following kits: GoPump kit 510080-BP, 5 fenestrated catheter & BIOPATCH; GoPump kit 510112-BP, 2.5" fenestrated catheter & BIOPATCH Go Pump Kit 510141-BP,10" fenestrated catheter& BIOPATCH Disposable Single Infusion Pump for drug delivery Symbios GOPump Rapid Recovery System is a local pain mangagement system that consists of a small balloon that is inflated with a local anesthetic medication.

Reason for Recall

Symbios Medical Products, Indianapolis, Indiana, is performing a voluntary recall of all GoPumps found in the GoPump Rapid Recovery System kits and all GoBlock kits manufactured with flow control components assembled prior to July 2012. The recall is being expanded after receiving additional reports relating to excessively high flow rates on the 150 mL reservoir, 2 mL/hr GOPump (PN510031), and the

Distribution Pattern

US Nationwide distribution.

Code Information

Pump 510076 Lot 11-100222 to 12-101087 found in: Kit 510080 BP. Lot 11-100215 to 12-100508 Kit 510141-BP Lot 12-100460 to 12-100673 Kit 510112-BP, Lot 11-100839-to 13-100225