Z-1463-2020 Class II Terminated

Recalled by RAYSEARCH LABORATORIES AB — Stockholm

FDA device recall Z-1463-2020 was initiated by RAYSEARCH LABORATORIES AB on Feb 20, 2020 and is designated Class II. Reason for recall: There is a potential for erroneous propagation of the current tabletop positions when applying pitch or roll correction. Even though pitch and roll correction is never propagated, non-zero pitch or… The recall status is terminated (terminated Nov 5, 2024). Affected quantity: 1 system in the U.S.

Recall details

Product Type
Devices
Report Date
Mar 18, 2020
Initiation Date
Feb 20, 2020
Termination Date
Nov 5, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1 system in the U.S.

Product description

RayStation/RayPlan 8A Service Pack 1, Build Number 8.0.1.10, stand-alone software treatment planning system

Reason for recall

There is a potential for erroneous propagation of the current tabletop positions when applying pitch or roll correction. Even though pitch and roll correction is never propagated, non-zero pitch or roll causes the lateral, longitudinal and vertical values propagated to be incorrect.

Distribution pattern

The products were distributed to the following US states: TN

Code information

UDI 07350002010136

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1463-2020
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for RAYSEARCH LABORATORIES AB? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for RAYSEARCH LABORATORIES AB: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.