Z-1469-2021 Class II Ongoing

Recalled by Heartware, Inc. — Miami Lakes, FL

FDA device recall Z-1469-2021 was initiated by Heartware, Inc. on February 26, 2021 and is designated Class II. Reason for recall: IFU and Patient Manual updated to describe Expected Alarm Duration. The Instructions for Use (IFU) is being updated to advise clinicians that alarms that resolve after a very short period may mean th… The recall status is ongoing. Affected quantity: 23123 (11292 US).

Recall Details

Product Type
Devices
Report Date
April 28, 2021
Initiation Date
February 26, 2021
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
23123 (11292 US)

Product Description

Instructions for Use and Patient Manual for HeartWare HVAD System. Accompanies these devices: Heartware HVAD pump implant kits Models 1100, 1101, 1102, 1103, 1104, 1104JP, 1205, and MCS1705PU. Heartware HVAD controller kits Models 1403US, 1407AU, 1407CA, 1407CH, 1407DE, 1407GB, 1407IL, 1407IN, 1407IT, 1407JP, 1407KR, 1420, and 1420JP. Heartware HVAD controller AC adapter Models 1430AR, 1430AU, 1430CA, 1430CH, 1430DE, 1430GB, 1430IL, 1430IN, 1430IT, 1430JP, and 1430US. Heartware HVAD controller DC Adapter Model 1440. Heartware HVAD data cable - monitor Model 1575. Heartware HVAD battery pack Models 1650, 1650CA-CLIN, and 1650DE. Heartware HVAD patient pack Model 1475. Heartware HVAD waist pack Models 2050, 2050IL, and 2050US. Heartware HVAD shoulder pack Models 2060, 2060IL, and 2060US.

Reason for Recall

IFU and Patient Manual updated to describe Expected Alarm Duration. The Instructions for Use (IFU) is being updated to advise clinicians that alarms that resolve after a very short period may mean there is an intermittent problem and the cause of the alarm can be evaluated by analysis of the controller log files. The Patient Manual (PM) is being updated to instruct patients to contact their clinicians if they cannot identify an alarm before it is cleared as it might mean there is an intermittent problem. NOTE: this update is included in the February 2021 Urgent Medical Device Notice also covered in recalls 87456 and 87604

Distribution Pattern

Worldwide distribution, including: US nationwide, Italy, Great Britain, Germany, Japan

Code Information

All serial numbers as of February 26, 2021. Model, GTIN: 1100, No GTIN Available; 1101, No GTIN Available; 1102, No GTIN Available; 1103, 00888707000017; 1104, No GTIN Available; 1104JP, 00888707000697; 1205, No GTIN Available; MCS1705PU, 00888707005364; 1403US, 00888707000475; 1407AU, 00888707001663; 1407CA, 00888707002851; 1407CH, 00888707001694; 1407DE, 00888707001700; 1407GB, 00888707001717; 1407IL, 00888707001724; 1407IN, 00888707001731; 1407IT, 00888707001748; 1407JP, 00888707001755; 1407KR, 00888707001762; 1420, 00888707000420; 1420JP, 00888707000437; 1430AR, 00888707000826; 1430AU, 00888707000833; 1430CA, 00888707000857; 1430CH, 00888707000871; 1430DE, 00888707000864; 1430GB, 00888707000888; 1430IL, 00888707000901; 1430IN, 00888707000505; 1430IT, 00888707000895; 1430JP, 00888707000918; 1430US, 00888707000307; 1440, 00888707001885; 1575, 00888707008051; 1650, 00888707000376; 1650CA-CLIN, 00888707001588; 1650DE, 00888707001373; 1475, 00888707008037; 2050, 00888707008082; 2050IL, 00888707001410; 2050US, 00888707000048; 2060, 00888707008099; 2060IL, 00888707001427; 2060US, 00888707000055;