Z-1477-2018 Class II Terminated

Recalled by Ion Beam Applications S.A. — Louvain La Neuve, N/A

Recall Details

Product Type
Devices
Report Date
May 2, 2018
Initiation Date
January 26, 2018
Termination Date
April 27, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5

Product Description

Proteus 235, ProteusPLUS or Proteus ONE produces and delivers a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.

Reason for Recall

When switching from Pencil Beam Scanning treatment mode to Uniform Scanning treatment mode, the tuning setpoints of the scanning magnets are not always taken into account when the proton beam is requested. The mismatch between the setpoint and feedback leads to an error message. Depending on user action, two scenarios are possible: 1. The user resumes the treatment field without analyzing the cause for the error message. This may lead to under-irradiation. 2. The user reboots the Scanning Magnets Power Supply Electronic Unit. This may lead to delay in patient treatment and the patient may need to be re-aligned, which leads to an additional X-ray dose.

Distribution Pattern

Consignees in four states: VA, IL, NJ, and WA. One consignee located in Germany.

Code Information

Affected version: PTS-7.X.X and PTS-8.X.X Serial numbers PAT.107 (EU); PAT.108, PAT.112, PAT.113, PAT.116 (US)