Z-1478-2026 Class II Ongoing
FDA device recall Z-1478-2026 was initiated by Olympus Corporation Of The Americas on Nov 14, 2025 and is designated Class II. Reason for recall: Ongoing investigations identified additional instances of the device Generator remaining in a blinking phase waiting to recognize the transducer. Damage to the transducer plug and/or generator… The recall status is ongoing. Affected quantity: 1082 units.
Recall details
- Product Type
- Devices
- Report Date
- Mar 11, 2026
- Initiation Date
- Nov 14, 2025
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1082 units
Product description
Brand Name: ShockPulse-SE Lithotripsy System SPL-S Product Name: ShockPulse-SE Lithotripsy System - Single Use Probes Model/Catalog Number: SPL-S Product Description: The ShockPulse Lithotripsy System is an electromechanical device capable of fragmenting calculi and aspirating stone debris. The ShockPulse transducer is a reusable device that requires cleaning and steam sterilization prior to use in order to generate both low-frequency mechanical impacts and high-frequency ultrasonic energy that quickly disintegrates stones.
Reason for recall
Ongoing investigations identified additional instances of the device Generator remaining in a blinking phase waiting to recognize the transducer. Damage to the transducer plug and/or generator receptacle may cause these issues.
Distribution pattern
Worldwide distribution - US Nationwide and the countries of Canada, Mexico, Bolivia, Chile, India, China, EU, Singapore, Australia, Korea, and Japan.
Code information
Model/Catalog Number: SPL-S; UDI: 00821925044197; Serial Numbers: All;
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1478-2026
- openFDA file date
- Nov 10, 2026
- Loaded by RegDataLab