Z-1481-2017 Class II Terminated

Recalled by Philips Medical Systems (Cleveland) Inc — Cleveland, OH

FDA device recall Z-1481-2017 was initiated by Philips Medical Systems (Cleveland) Inc on Jan 31, 2017 and is designated Class II. Reason for recall: Four issues: 1. Motion controller problem stops scan and no data image produced. 2. Door interlock switch problem disables CT scan. 3. Detector contacts head holder when performing Patient… The recall status is terminated (terminated Jul 2, 2018). Affected quantity: 1218 total.

Recall details

Product Type
Devices
Report Date
Mar 22, 2017
Initiation Date
Jan 31, 2017
Termination Date
Jul 2, 2018
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1218 total

Product description

882478: BrightView X designed for single or dual detector nuclear imaging accommodating a range of ECT studies. In addition it can be used to perform planar static, dynamic, gated, total body, circular-orbit and noncircular orbit SPECT, gated SPECT (circular and noncircular)studies, computer-programmed protocol strings, and reference scans (dual detectors).

Reason for recall

Four issues: 1. Motion controller problem stops scan and no data image produced. 2. Door interlock switch problem disables CT scan. 3. Detector contacts head holder when performing Patient Unload. 4. JETStream freezes during gated planar scan

Distribution pattern

Worldwide distribution. US nationwide, ALGERIA, ARGENTINA, ARMENIA, AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, BRAZIL, CANADA, CAYMEN ISLANDS, CHILE, CHINA, CZECH REPUBLIC, DENMARK, DOMINICAN REPUBLIC, EGYPT, FINLAND, FRANCE, GABON, GERMANY, GREAT BRITAIN, GREECE, INDIA, INDONESIA, IRAN, IRELAND, ISRAEL, ITALY, JAPAN, KOREA, LEBANON, LITHUANIA, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, OMAN, PALESTINE, PANAMA, PERU, PHILIPPINES, POLAND, PORTUGAL, RUSSIAN FEDERATION, SAUDI ARABIA, SINGAPORE, SLOVENIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, SYRIAN ARAB REPUBLIC, TAIWAN, THAILAND, TURKEY, UKRAINE, UNITED ARAB EMIRATES, UNITED KINGDOM, & VIETNAM.

Code information

11000100, 11000019, 11000034, 11000020, 11000112, 11000086, 11000045, 11000092, 11000037, 11000038, 11000106, 11000059, 11000064, 11000039, 11000087, 11000004, 11000058, 11000030, 11340005, 11000113, 11000117, 11000011, 11000118, 11000074, 11000075, 11000036, 11000104, 11000103, 11000071, 11340002, 11000089, 11000012, 11000041, 11000042, 11000043, 11000049, 11000014, 11000099, 11000022, 11000031, 11000005, 11000028, 11000073, 11000040, 11000077, 11000008, 11000048, 11000093, 11000096, 11000097, 11000069, 11000079, 11000080, 11000024, 11000076, 11000085, 11000108, 11000109, 11000115, 11000056, 11000029, 11340003, 11000044, 11000107, 11000090, 11000017, 11000110, 11000009, 11000035, 11000081, 11000062, 11000013, 11000015, 11000063, 11000083, 11000091, 11000098, 11000054, 11340004, 11000102, 11000061, 11000095, 11000023, 11000094, 11000060, 11000055, 11000070, 11000084, 11000088, 11000068, 11000052, 11000057, 11000050, 11000067, 11000007, 11000010, 11000027, 11000032, 11340001, 11000026, 11000114, 11000078, 11000051, 11000046, 11000003, 11000018, 11000066, 11000047, 11000101, 11000065, 11000072, 11000116, 11000016, 11000105, 11000025, 11000006, 11000033,

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1481-2017
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Philips Medical Systems (Cleveland) Inc? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Philips Medical Systems (Cleveland) Inc: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Philips Medical Systems (Cleveland) Inc →

Source: openFDA record for Z-1481-2017. Report a data issue: hello@regdatalab.com.