Z-1484-2018 Class I Terminated

Recalled by Draeger Medical, Inc. — Telford, PA

Recall Details

Product Type
Devices
Report Date
May 16, 2018
Initiation Date
March 1, 2018
Termination Date
April 16, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1

Product Description

Fabius Tiro M, anesthesia machine, catalog no. 8608400 Product Usage: The products are inhalation anesthesia machines for use in operating, induction and recovery rooms. They may be used with O2, N2O, and AIR supplied by a medical gas pipeline system or by externally mounted gas cylinders. Fabius series anesthesia machines are equipped with an electrically driven and electronically controlled ventilator and monitors for airway pressure (P), volume (V), and inspiratory oxygen concentration (FiO2).

Reason for Recall

The automatic ventilation may fail if the position detection of the motor is disturbed. If the ventilator cover is damaged, manual ventilation may also fail.

Distribution Pattern

US Nationwide Distribution to accounts in 21 states: AZ, CA, CO, ID, IL, KS, MA, MI, MS, NC, NY, OH, OK, PA, RI, TN, TX, UT, VA, WA, WI; and PR.

Code Information

Serial no. ASKK-0001