Z-1496-2017 Class II Terminated
FDA device recall Z-1496-2017 was initiated by Merge Healthcare, Inc. on Sep 22, 2015 and is designated Class II. Reason for recall: For sites using the Issuer of Patient ID (IPID), the system will display the study list and images from different patients with different IPIDs as though they are for the same patient if they all have… The recall status is terminated (terminated Apr 18, 2017). Affected quantity: 16 sites were originally identified as potentially having the affected versions.
Recall details
- Product Type
- Devices
- Report Date
- Mar 29, 2017
- Initiation Date
- Sep 22, 2015
- Termination Date
- Apr 18, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 16 sites were originally identified as potentially having the affected versions
Product description
Merge Cardio software with Issuer of Patient ID (IPID). The firm name on the label is Merge Healthcare, Hartland, WI.
Reason for recall
For sites using the Issuer of Patient ID (IPID), the system will display the study list and images from different patients with different IPIDs as though they are for the same patient if they all have the first name, last name, and medical record number in common.
Distribution pattern
US Distribution was made to medical facilities located in GA, IL, MD, MI, MO, OH, TN, and VT. There was no foreign/government/military distribution.
Code information
Versions 8.30, 9.0, 9.0.1, 9.0.2, 9.0.3, 9.0.4, 9.0.5, 9.0.6, 9.0.7, 9.0.8, 10.0, 10.0.1, and 10.1 with Issuer of Patient ID (IPID)