Z-1497-2015 Class II Terminated
FDA device recall Z-1497-2015 was initiated by Wright Medical Technology, Inc. on Mar 6, 2015 and is designated Class II. Reason for recall: The minor diameter, major diameter, thread pitch, and pitch on the Crosscheck are out of specification, which may result in intraoperative malfunction The recall status is terminated (terminated Aug 5, 2018). Affected quantity: 138 units.
Recall details
- Product Type
- Devices
- Report Date
- Apr 29, 2015
- Initiation Date
- Mar 6, 2015
- Termination Date
- Aug 5, 2018
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 138 units
Product description
Reason for recall
Distribution pattern
Nationwide Distribution.