Z-1509-2015 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- April 29, 2015
- Initiation Date
- March 9, 2015
- Termination Date
- January 12, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 14,428 Universal Nuts
Product Description
TS Crown Component 4UR (5/pk), TS Crown Component 5UR (5/pk), TS Crown Component 6UR (5/pk), TS Crown Component 4LR (5/pk), TS Crown Component 5LR (5/pk), TS Crown Component 6LR (5/pk), TS Crown Component 4UL (5/pk), TS Crown Component 5UL (5/pk), TS Crown Component 4LL (5/pk), TS Crown Component 5LL (5/pk), and TS Crown Component 6LL (5/pk) A Herbst Appliance is used to affect changes in the skeletal relationship and/or occlusion of a patient. the Universal Nut is a component of a Herbst Appliance
Reason for Recall
The Universal Nut, a component of the orthodontic Herbst Appliance was made with 416 SS and not the specified 303 SS.
Distribution Pattern
Worldwide Distribution - US nationwide including Puerto Rico, Brazil, China, Colombia, Germany, Ireland, Mexico, Switzerland, and United Kingdom.
Code Information
Product No. 91672, 91673, 91674, 91677, 91678, 91679, 91682, 91683, 91687, 91688, and 91689