Z-1511-2015 Class II Terminated

Recalled by Specialty Appliance Works, Inc. — Cumming, GA

Recall Details

Product Type
Devices
Report Date
April 29, 2015
Initiation Date
March 9, 2015
Termination Date
January 12, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
14,428 Universal Nuts

Product Description

TS ROC Crown Component 4UR (5/pk), TS ROC Crown Component 5UR (5/pk), TS ROC Crown Component 6UR (5/pk), TS ROC Crown Component 7UR (5/pk), TS ROC Crown Component 8UR (5/pk), TS ROC Crown Component 4LR (5/pk), TS ROC Crown Component 5LR (5/pk), TS ROC Crown Component 6LR (5/pk), TS ROC Crown Component 7LR (5/pk), TS ROC Crown Component 8LR (5/pk), TS ROC Crown Component 4UL (5/pk), TS ROC Crown Component 5UL (5/pk), TS ROC Crown Component 6UL (5/pk), TS ROC Crown Component 7UL (5/pk), TS ROC Crown Component 8UL (5/pk), TS ROC Crown Component 4LL (5/pk), TS ROC Crown Component 5LL (5/pk), TS ROC Crown Component 6LL (5/pk), TS ROC Crown Component 7LL (5/pk), TS ROC Crown Component 8LL (5/pk) A Herbst Appliance is used to affect changes in the skeletal relationship and/or occlusion of a patient. the Universal Nut is a component of a Herbst Appliance.

Reason for Recall

The Universal Nut, a component of the orthodontic Herbst Appliance was made with 416 SS and not the specified 303 SS.

Distribution Pattern

Worldwide Distribution - US nationwide including Puerto Rico, Brazil, China, Colombia, Germany, Ireland, Mexico, Switzerland, and United Kingdom.

Code Information

Product No. 91720, 91721, 91722, 91723, 91724, 91725, 91726, 91727, 91728, 91729, 91730, 91731, 91732, 91733, 91734, 91735, 91736, 91737, 91738, and 91739