Z-1528-2013 Class II Terminated
FDA device recall Z-1528-2013 was initiated by Bard Peripheral Vascular Inc on Jan 23, 2013 and is designated Class II. Reason for recall: Bard Peripheral Vascular (BPV) is initiating this recall because a combination of Bard¿ UltraClip¿ Dual Trigger Tissue Marker's and support catheters (Microsheath and Usher) were inadvertently… The recall status is terminated (terminated Jun 10, 2013). Affected quantity: 17.
Recall Details
- Product Type
- Devices
- Report Date
- Jun 19, 2013
- Initiation Date
- Jan 23, 2013
- Termination Date
- Jun 10, 2013
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 17
Product Description
MICROSHEATH¿ - 0.066 (1.7mm) proximal, 0.064 (1.6mm); WL 123cm, RGC 8F Guide; minimum ID: 0.086"/ (2.2mm);distal, straight non-tapered tip; PK1018-03A; Bard Peripheral Vascular; Manuf: Flow Cardia, Inc., a Subsidiary of C.R. Bard, Inc. 745 North Pastoria Ave., Sunnyvale, CA 94085. The MicroSheath and Usher Support Catheters are single lumen catheters intended to create a pathway for other devices in the peripheral vasculature. MicroSheath¿ is a single lumen catheter that has a 123cm working length and a straight tip shape
Reason for Recall
Bard Peripheral Vascular (BPV) is initiating this recall because a combination of Bard¿ UltraClip¿ Dual Trigger Tissue Marker's and support catheters (Microsheath and Usher) were inadvertently distributed to customers without completing the sterilization process (non sterile).
Distribution Pattern
Nationwide Distribution including AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, IL, IN, KY, KS, LA, MA, MD, MI, MN, MO, MS, NC, NJ, NY, OH, OK, PA, TX, VA, WA, WV and WY.
Code Information
PC MS17123; Lot# FCWC10010
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1528-2013
- openFDA file date
- Oct 23, 2026
- Loaded by RegDataLab