Z-1537-2014 Class II Terminated

Recalled by Bio-Rad Laboratories, Inc. — Hercules, CA

FDA device recall Z-1537-2014 was initiated by Bio-Rad Laboratories, Inc. on Apr 8, 2014 and is designated Class II. Reason for recall: There is an issue with the buffers which impacts the quantification of the Hemoglobin F peak. Calibration and Quality control are likely to fail if the affected lots are used. The recall status is terminated (terminated May 30, 2014). Affected quantity: 457 kits.

Recall details

Product Type
Devices
Report Date
May 14, 2014
Initiation Date
Apr 8, 2014
Termination Date
May 30, 2014
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
457 kits

Product description

Variant II beta-thalassemia Short Program Model Numbers 270-2103 (250 tests) and 270-2154 (500 tests), Bio-Rad Laboratories, Inc. in vitro diagnostic. The Variant II Hemoglobin testing system uses the principals of high performance liquid chromatography (HPLC) for the separation and determination of normal and abnormal hemoglobin.

Reason for recall

There is an issue with the buffers which impacts the quantification of the Hemoglobin F peak. Calibration and Quality control are likely to fail if the affected lots are used.

Distribution pattern

Worldwide Distrubution ; USA INCLUDING CA, NY, GA, NC, AL, NJ, CT, WA, NY, LA, Il, MA and Internationally to China, Australia, Canada, France, Great Britain, Hong Kong, Thailand, Singapore,and Brazil.

Code information

Catalog number: 270-2103 (250 tests; Reorder pack lot number: 64005302, Expiration: 4/2015; Catalog number: 270-2154 (500 tests) Reorder pack lot number: 64005207, Expiration: 6/2015; 64004750, Expiration: 11/2015

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1537-2014
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Bio-Rad Laboratories, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Bio-Rad Laboratories, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Bio-Rad Laboratories, Inc. →

Source: openFDA record for Z-1537-2014. Report a data issue: hello@regdatalab.com.