Z-1544-2013 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- June 26, 2013
- Initiation Date
- January 17, 2013
- Termination Date
- November 7, 2013
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3615 units total
Product Description
SybronEndo Elements Obturation Unit; REF 973-0322; SN 91-1234; 2005-10; SybronEndo, 1332 S. Lone Hill Ave. Glendora, CA 91740; EU Rep: Kerr Italia, S.r.l, Via Passanti, 332, 1-84018 Scafati (SA), Italy The intended use of the Elements Obturation Unit and Replacement A/C Power Cord is to provide continuous heat at the tip of the dental instrument to test tooth response to thermal stimulus, for tissue cauterization and coagulation and to backfill and down pack gutta percha during endodontic root canal treatment.
Reason for Recall
A recall has been issued by SybronEndo because a Canadian Standards Association (CSA)-directed cautionary statement label was inadvertently omitted from the Elements Obturation Unit and its Replacement A/C Power Cord.
Distribution Pattern
Worldwide Distribution - USA (nationwide) and the countries of Australia, Azerbaijan, Belgium, Canada, China, Cyprus, Czech Republic, Egypt, Estonia, Finland, France, Germany, Greece, Hong Kong, Iceland, India, Italy, Japan, Kuwait, Latvia, Lebanon, Malaysia, Netherlands, New Zealand, Pakistan, Poland, Romania, Russia, Saudi Arabia, Singapore, Spain, Sweden, Switzerland, Thailand, Turkey, Ukraine, United Arab Emirates, and United Kingdom.
Code Information
Lot/Serial Numbers: 91-17011 to 91-19095; 91-20889 to 91-22339