Z-1550-2015 Class II Terminated
FDA device recall Z-1550-2015 was initiated by Fresenius Medical Care Holdings, Inc. on Feb 20, 2015 and is designated Class II. Reason for recall: The Acetate value for GranuFlo on the Select Concentrate screen should be 8mEq/l but it is displayed as 4mEq/l. The correct value is shown on the dialysate screen. The recall status is terminated (terminated May 18, 2021). Affected quantity: 133,751 in total (US: 130,337 and OUS: Canada 1,201 and Mexico 2,213).
Recall details
- Product Type
- Devices
- Report Date
- May 6, 2015
- Initiation Date
- Feb 20, 2015
- Termination Date
- May 18, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 133,751 in total (US: 130,337 and OUS: Canada 1,201 and Mexico 2,213)
Product description
2008 Series Hemodialysis Machines: 2008T with the following product code and description: 190395 2008K at Home System - 1st Gen. -¿ Canada; 190828 2008K at Home System - 2nd Gen - U.S.; 190904 2008K at Home System w/ Bibag; 190828 2008KatHOME HEMODIALYSIS SYSTEM; 190395 2008KatHOME MACHINE, SHORT CAB, OLC/DP, HP The Fresenius 2008K is indicated for acute and chronic dialysis therapy.
Reason for recall
The Acetate value for GranuFlo on the Select Concentrate screen should be 8mEq/l but it is displayed as 4mEq/l. The correct value is shown on the dialysate screen.
Distribution pattern
Worldwide Distribution: US (Nationwide) and countries of: Canada and Mexico.
Code information
Serial Numbers: 0T0S100031- 101358, 1T0S101359- 127938, 2T0S109489- 118130, 3T0S118131- 126163, 4T0S126164- 135408, 9T0S100002, 9T0S100004, 9T0S100007- 100010, 9T0S100012, 9T0S100014- 100016
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1550-2015
- openFDA file date
- Nov 16, 2026
- Loaded by RegDataLab
- Nov 17, 2026