Z-1570-2016 Class II Terminated
FDA device recall Z-1570-2016 was initiated by Becton, Dickinson And Company, BD Biosciences on Mar 31, 2016 and is designated Class II. Reason for recall: One lot of lot of BD Trucount tubes may contain tubes from another lot. There is a 3.3% difference in bead counts between the two lots found within some of the pouches and presents a low risk to… The recall status is terminated (terminated May 26, 2017). Affected quantity: 1743; 467 US and 1276 International.
Recall details
- Product Type
- Devices
- Report Date
- May 4, 2016
- Initiation Date
- Mar 31, 2016
- Termination Date
- May 26, 2017
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1743; 467 US and 1276 International
Product description
BD Trucount Absolute Counting Tubes: Model number: 340334 IVD; BD Trucount tubes are used for determining absolute counts of leucocytes in blood. BD Trucount tubes are designed for use with in vitro diagnostic products such as BD Tritest reagents, and a suitably equipped flow cytometer. BD Trucount tubes can be used with the BD FACS Loader.
Reason for recall
One lot of lot of BD Trucount tubes may contain tubes from another lot. There is a 3.3% difference in bead counts between the two lots found within some of the pouches and presents a low risk to patient safety or impact to data if this product is used.
Distribution pattern
Worldwide Distribution - US including MD, CA GA. WA. PA. CO. TN. FL. NC. DE. VA, CT, LA. NC, NY, OH, MN, NH and Internationally to China, Japan, and India.
Code information
Model: 340334; Manufacaturing lot: 534964