Z-1570-2016 Class II Terminated

Recalled by Becton, Dickinson And Company, BD Biosciences — San Jose, CA

FDA device recall Z-1570-2016 was initiated by Becton, Dickinson And Company, BD Biosciences on Mar 31, 2016 and is designated Class II. Reason for recall: One lot of lot of BD Trucount tubes may contain tubes from another lot. There is a 3.3% difference in bead counts between the two lots found within some of the pouches and presents a low risk to… The recall status is terminated (terminated May 26, 2017). Affected quantity: 1743; 467 US and 1276 International.

Recall details

Product Type
Devices
Report Date
May 4, 2016
Initiation Date
Mar 31, 2016
Termination Date
May 26, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1743; 467 US and 1276 International

Product description

BD Trucount Absolute Counting Tubes: Model number: 340334 IVD; BD Trucount tubes are used for determining absolute counts of leucocytes in blood. BD Trucount tubes are designed for use with in vitro diagnostic products such as BD Tritest reagents, and a suitably equipped flow cytometer. BD Trucount tubes can be used with the BD FACS Loader.

Reason for recall

One lot of lot of BD Trucount tubes may contain tubes from another lot. There is a 3.3% difference in bead counts between the two lots found within some of the pouches and presents a low risk to patient safety or impact to data if this product is used.

Distribution pattern

Worldwide Distribution - US including MD, CA GA. WA. PA. CO. TN. FL. NC. DE. VA, CT, LA. NC, NY, OH, MN, NH and Internationally to China, Japan, and India.

Code information

Model: 340334; Manufacaturing lot: 534964

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1570-2016
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1570-2016. Report a data issue: hello@regdatalab.com.