Z-1572-2026 Class II Ongoing
FDA device recall Z-1572-2026 was initiated by Howmedica Osteonics Corp. on Feb 2, 2026 and is designated Class II. Reason for recall: A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen… The recall status is ongoing. Affected quantity: 1098 units.
Recall details
- Product Type
- Devices
- Report Date
- Mar 25, 2026
- Initiation Date
- Feb 2, 2026
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1098 units
Product description
Reason for recall
Distribution pattern
Domestic: Nationwide Distribution; International: Australia, Germany, Belgium, France, Spain UK, New Zealand;
Code information
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1572-2026
- openFDA file date
- Nov 24, 2026
- Loaded by RegDataLab
- Nov 25, 2026