Z-1572-2026 Class II Ongoing

Recalled by Howmedica Osteonics Corp. — Mahwah, NJ

FDA device recall Z-1572-2026 was initiated by Howmedica Osteonics Corp. on February 2, 2026 and is designated Class II. Reason for recall: A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen… The recall status is ongoing. Affected quantity: 1098 units.

Recall Details

Product Type
Devices
Report Date
March 25, 2026
Initiation Date
February 2, 2026
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1098 units

Product Description

MICS3 Angled Sagittal Saw Attachment; Part Number: 210490

Reason for Recall

A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.

Distribution Pattern

Domestic: Nationwide Distribution; International: Australia, Germany, Belgium, France, Spain UK, New Zealand;

Code Information

Part Number: 210490; UDI: (01)07613327523874(10)TU6563205RP; Lot numbers: ALL;