Z-1597-2021 Class II Terminated

Recalled by RAYSEARCH LABORATORIES AB — Stockholm, N/A

FDA device recall Z-1597-2021 was initiated by RAYSEARCH LABORATORIES AB on Feb 22, 2021 and is designated Class II. Reason for recall: For some LINAC types, merging clinical beams with beams of approximate dose may lead to the approximate dose erroneously labeled as clinical dose. Merge beams can be used manually, in scripting, or as… The recall status is terminated (terminated May 1, 2023). Affected quantity: 116 units.

Recall details

Product Type
Devices
Report Date
May 26, 2021
Initiation Date
Feb 22, 2021
Termination Date
May 1, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
116 units

Product description

RayStation 4.5, 4.7, 4.9, 5, 6, 7, 8A, 8B, 9A, 9B, 10A, 10B, RayPlan 1, 2, 7, 8A, 8B, 9A, 10A, 10B including some of the Service Packs. Standalone software treatment planning system

Reason for recall

For some LINAC types, merging clinical beams with beams of approximate dose may lead to the approximate dose erroneously labeled as clinical dose. Merge beams can be used manually, in scripting, or as part of the Automated breast planning feature. In some cases, when merging two beams where the first beam has clinical dose and the second beam has approximate dose, the dose of the resulting beam will be labeled as Clinical, although dose for some of the beam segments is still calculated with the SVD dose engine and it should be labeled Approximate: Mixed dose. The difference between approximate and final Clinical dose is in most cases small, but there can be body sites such as lung where the difference can be significant.

Distribution pattern

US Nationwide distribution.

Code information

Build number: 4.5.1.14 , 4.7.2.5 , 4.7.3.13 , 4.7.4.4 , 4.7.5.4 , 4.7.6.7 , 4.9.0.42 , 5.0.1.11 , 5.0.2.35 , 5.0.3.17 , 6.0.0.24 , 6.1.1.2 , 6.2.0.7 , 6.3.0.6 , 7.0.0.19 , 8.0.0.61 , 8.0.1.10 , 8.1.0.47 , 8.1.1.8 , 8.1.2.5 , 9.0.0.113 , 9.1.0.933 , 9.2.0.483 , 10.0.0.1154 , 10.0.1.52 or 10.1.0.613 UDI: RayStation 4.0 (4.0.0.14) to RayStation 5 Service Pack 2 (5.0.2.35) N/A RayStation 5 Service Pack 3 (5.0.3.17) UDI 07350002010020 RayStation 6/RayPlan 2 (6.0.0.24) UDI 07350002010013 RayStation 6/RayPlan 2 Service Pack 1 (6.1.1.2) UDI 07350002010082 RayStation 6/RayPlan 2 Service Pack 2 (6.2.0.7) UDI 07350002010075 RayStation 6/RayPlan 2 Service Pack 3 (6.3.0.6) UDI 07350002010242 RayStation/RayPlan 7 (7.0.0.19) UDI 07350002010068 RayStation/RayPlan 8A (8.0.0.61) UDI 07350002010112 RayStation/RayPlan 8A Service Pack 1 (8.0.1.10) UDI 07350002010136 RayStation/RayPlan 8B (8.1.0.47) UDI 07350002010129 RayStation/RayPlan 8B Service Pack 1 (8.1.1.8) UDI 07350002010204 RayStation/RayPlan 8B Service Pack 2 (8.1.2.5) UDI 07350002010235 RayStation/RayPlan 9A (9.0.0.113) UDI 07350002010174 RayStation/RayPlan 9B (9.1.0.933) UDI 07350002010266 RayStation/RayPlan 9B Service Pack 1 (9.2.0.483) UDI 07350002010297 RayStation/RayPlan 10A (10.0.0.1154) UDI 07350002010303 RayStation/RayPlan 10A Service Pack 1 (10.0.1.52) UDI 07350002010365 RayStation/RayPlan 10B (10.1.0.613) UDI 07350002010310

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1597-2021
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for RAYSEARCH LABORATORIES AB? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for RAYSEARCH LABORATORIES AB: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about RAYSEARCH LABORATORIES AB →

Source: openFDA record for Z-1597-2021. Report a data issue: hello@regdatalab.com.