Z-1604-2026 Class II Ongoing

Recalled by I.T.S. GmbH — Lasnitzhohe

FDA device recall Z-1604-2026 was initiated by I.T.S. GmbH on Jan 23, 2026 and is designated Class II. Reason for recall: Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU). The recall status is ongoing. Affected quantity: 120 units.

Recall details

Product Type
Devices
Report Date
Apr 1, 2026
Initiation Date
Jan 23, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
120 units

Product description

I.T.S. DUL (Distal Ulna Locking) System and UOL (Ulna Osteotomy Locking) System with the below descriptions and corresponding article numbers. 1. I.T.S. DUL (Distal Ulna Locking) System, Distal Ulna Plate; Article Numbers: 21232-3, 21232-4, 21232-6, 21233-3, 21233-4, 21233-6, 21234-4, 21235-4. 2. I.T.S. UOL (Ulna Osteotomy Locking) System, Ulna Osteotomy Plate; Article Numbers: 21231-6. 3. UOL (Ulna Osteotomy Locking) System, Cancellous Screw, Locking; Article Numbers: 37302-18, 37302-20.

Reason for recall

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Distribution pattern

US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.

Code information

1. Article Number (Lot Numbers): 21232-3 (26/F65636), 21232-4 (26/F61722, 26/F65637), 21232-6 (26/F62564, 26/F65638), 21233-3 (26/F65639), 21233-4 (26/F65640), 21233-6 (26/F61723, 26/F65046), 21234-4 (26/F65641, 26/F66504), 21235-4 (26/F65642, 26/F67810). 2. Article Number (Lot Numbers): 21231-6 (643/012301, 643/022310, 643/032308, 643/0630, 643/0657, 643/102402, 643/122301). 3. Article Number (Lot Numbers): 37302-18 (18/636447), 37302-20 (18/802189).

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1604-2026
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1604-2026. Report a data issue: hello@regdatalab.com.