Z-1605-2024 Class II Ongoing

Recalled by SEASPINE ORTHOPEDICS CORPORATION — Carlsbad, CA

FDA device recall Z-1605-2024 was initiated by SEASPINE ORTHOPEDICS CORPORATION on Feb 15, 2024 and is designated Class II. Reason for recall: Screw heads, of pedicle screw systems, have thin lead-in thread that pose a risk of intraoperative cross-threading within the locking cap/screw head interface. The recall status is ongoing. Affected quantity: 1579 Screw Heads.

Recall details

Product Type
Devices
Report Date
May 1, 2024
Initiation Date
Feb 15, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1579 Screw Heads

Product description

Mariner RDX Polyaxial Head, REF: MX1-000020; Mariner RDX Extended Polyaxial Head, REF: MX1-000030; Mariner RDX Deformity Head, REF: MX1-000040; Mariner RDX Extended Deformity Head, REF: MX1-000050; Mariner RDX Trauma Head, REF: MX1-000060; Mariner RDX Extended Trauma Head. REF: MX1-000070, part of Mariner RDX Alpha Instruments and Implants, REF: AMRDX and Mariner RDX TD Alpha Implants, REF: AMRDXTD, which are pedicle screw systems.

Reason for recall

Screw heads, of pedicle screw systems, have thin lead-in thread that pose a risk of intraoperative cross-threading within the locking cap/screw head interface.

Distribution pattern

US Nationwide distribution in the states of MI, GA, VA, IN, FL.

Code information

REF/UDI-DI/Lots: MX1-000020/10889981289751/AW161073C, AW164798D, AW164799D, AW164800E, AW164801E; MX1-000030/10889981289768/AW164802F, AW164803G, AW157588F, AW161074F, AW163331F; MX1-000040/10889981289775/AW157589D, AW164806E; MX1-000050/10889981289782/AW164807G, AW164808G, AW164809H, AW157590G, AW172902H; MX1-000060/10889981289799/AW157591D, AW161076E, AW164811E; MX1-000070/10889981289805/AW163350G, AW164812G, AW157592G, AW161077G; AMRDX/10889981291266/AMRDX-001, AMRDX-002, AMRDX-003, AMRDX-004, AMRDX-005, AMRDX-008, AMRDX-009; AMRDXTD/10889981295332/AMRDXTD-003, AMRDXTD-004, AMRDXTD-005

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1605-2024
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1605-2024. Report a data issue: hello@regdatalab.com.