Z-1620-2024 Class II Ongoing

Recalled by SEASPINE ORTHOPEDICS CORPORATION — Carlsbad, CA

FDA device recall Z-1620-2024 was initiated by SEASPINE ORTHOPEDICS CORPORATION on Feb 28, 2024 and is designated Class II. Reason for recall: Screw heads, of pedicle screw systems, have partial thread within the screw head that pose a risk of intraoperative cross-threading within the locking cap/screw head interface. The recall status is ongoing. Affected quantity: 444 Screw Heads.

Recall details

Product Type
Devices
Report Date
May 1, 2024
Initiation Date
Feb 28, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
444 Screw Heads

Product description

Mariner RDX MIS Polyaxial Head, REF: MX1-001010, part of Mariner RDX MIS Alpha Instruments and Implants, REF: AMRDXMIS, which are pedicle screw systems.

Reason for recall

Screw heads, of pedicle screw systems, have partial thread within the screw head that pose a risk of intraoperative cross-threading within the locking cap/screw head interface.

Distribution pattern

US: CA, FL, CO, TX

Code information

REF/UDI-DI/Lots: MX1-001010/10889981298647/ AW161319D, AW164195D, AW166994D, AW167309D, AW167422D, AW169350F, AW169351F. AMRDXMIS/10889981303297/ AMRDXMIS-001, AMRDXMIS-002, AMRDXMIS-003, AMRDXMIS-004, AMRDXMIS-005

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1620-2024
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1620-2024. Report a data issue: hello@regdatalab.com.