Z-1613-2023 Class II Ongoing

Recalled by Olympus Corporation Of The Americas — Center Valley, PA

FDA device recall Z-1613-2023 was initiated by Olympus Corporation Of The Americas on Apr 18, 2023 and is designated Class II. Reason for recall: The Single Use Distal Cover MAJ-2315 may cause mucosal injury and/or become detached from the endoscope during use. Consequences of a detached cover could include the risk of aspiration, inhalation… The recall status is ongoing. Affected quantity: 309278 pieces.

Recall details

Product Type
Devices
Report Date
May 31, 2023
Initiation Date
Apr 18, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
309278 pieces

Product description

Olympus Single Use Distal Cover Model Number MAJ-2315 used with EVIS EXERA III Duodenovideoscope TJF-Q190V; Item code: Catalogue #N5786100

Reason for recall

The Single Use Distal Cover MAJ-2315 may cause mucosal injury and/or become detached from the endoscope during use. Consequences of a detached cover could include the risk of aspiration, inhalation, or obstruction that will require urgent removal, and/or burns from an uncovered distal end.

Distribution pattern

US Nationwide.

Code information

UDI-DI: 04953170403019; Item code: Catalog N5786100 Lots Lower than H2530

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1613-2023
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Olympus Corporation Of The Americas? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Olympus Corporation Of The Americas: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Olympus Corporation Of The Americas →

Source: openFDA record for Z-1613-2023. Report a data issue: hello@regdatalab.com.