Z-1619-2014 Class II Terminated
FDA device recall Z-1619-2014 was initiated by Fresenius Medical Care Holdings, Inc. on Apr 25, 2014 and is designated Class II. Reason for recall: 2008-Series Hemodialysis Machines with Revision K Actuator Test Boards may fail to detect a disconnected Ultrafiltration (UF) Pump The recall status is terminated (terminated May 18, 2021). Affected quantity: 39 units.
Recall details
- Product Type
- Devices
- Report Date
- May 21, 2014
- Initiation Date
- Apr 25, 2014
- Termination Date
- May 18, 2021
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 39 units
Product description
Revision K Actuator Test Boards sold as replacement parts for use with Fresenius 2008-Series Hemodialysis Machines (2008K, 2008K2, 2008K@home and 2008T). Part Number: 190713 Revision K
Reason for recall
2008-Series Hemodialysis Machines with Revision K Actuator Test Boards may fail to detect a disconnected Ultrafiltration (UF) Pump
Distribution pattern
USA (nationwide) and the country of Canada.*
Code information
Revision K Actuator Test Boards
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-1619-2014
- openFDA file date
- Nov 16, 2026
- Loaded by RegDataLab