Z-1632-2026 Class II Ongoing

Recalled by Straumann USA LLC — Andover, MA

Recall Details

Product Type
Devices
Report Date
April 1, 2026
Initiation Date
March 17, 2026
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
13 units

Product Description

Custom Abutment Ti, for Straumann RC, incl. TAN screw; Catalog Number/Art.: 027.4620;

Reason for Recall

A material mix-up had occurred during the manufacturing of Straumann¿ Custom Abutments. More specifically, a raw material bar of TAN-material (SAP #701924 / Lot 7130439) was incorrectly identified as a Titanium Bar Ti Gr4 D (SAP #701922 / Lot 7130430). As a result, RC Titanium abutments were manufactured using the TAN Bar material on Mill-from-Bar Machine 396-06.

Distribution Pattern

Worldwide - US Nationwide distribution in the states of CA, CO, FL, IA, IL, MN, MO, MT, OR, WA; and the country of Canada.

Code Information

Catalog Number/Art.: 027.4620; UDI-DI: 07630031713766; Order number (lot): 92465US_12039_00_b00 , 92465US_12040_00_b00 , 92465US_12047_00_b00 , 92465US_12048_00_b00 , 70959US_02227_00_b00 , 94021US_02348_00_b00, 70416US_07543_00_b00 , 91395US_01687_00_b00 , 91545US_03284_00_b00 , 91545US_03285_00_b00 , 71711US_00420_00_b00 , 91853US_06313_00_b00 , 72747US_00247_00_b00;